Use of Virtual Reality to Assess Dynamic Posturography and Sensory Organization: Instrument Validation Study

Use of Virtual Reality to Assess Dynamic Posturography and Sensory Organization: Instrument Validation Study

Use of Virtual Reality to Assess Dynamic Posturography and Sensory Organization: Instrument Validation Study

Original Paper

1Department of Exercise Science, Elon University, Elon, NC, United States

2Department of Physics, Elon University, Elon, NC, United States

*all authors contributed equally

Corresponding Author:

Matthew William Wittstein, BSc, MSc, PhD

Department of Exercise Science

Elon University

2525 Campus Box

Elon, NC, 27244

United States

Phone: 1 336 278 6693

Email: mwittstein@elon.edu


Background: The Equitest system (Neurocom) is a computerized dynamic posturography device used by health care providers and clinical researchers to safely test an individual’s postural control. While the Equitest system has evaluative and rehabilitative value, it may be limited owing to its cost, lack of portability, and reliance on only sagittal plane movements. Virtual reality (VR) provides an opportunity to reduce these limitations by providing more mobile and cost-effective tools while also observing a wider array of postural characteristics.

Objective: This study aimed to test the plausibility of using VR as a feasible alternative to the Equitest system for conducting a sensory organization test.

Methods: A convenience sample of 20 college-aged healthy individuals participated in the study. Participants completed the sensory organization test using the Equitest system as well as using a VR environment while standing atop a force plate (Bertec Inc). The Equitest system measures the equilibrium index. During VR trials, the estimated equilibrium index, 95% ellipse area, path length, and anterior-posterior detrended fluctuation analysis scaling exponent alpha were calculated from center of pressure data. Pearson correlation coefficients were used to assess the relationship between the equilibrium index and center of pressure–derived balance measures. Intraclass correlations for absolute agreement and consistency were calculated to compare the equilibrium index and estimated equilibrium index.

Results: Intraclass correlations demonstrated moderate consistency and absolute agreement (0.5 < intraclass correlation coefficient < 0.75) between the equilibrium index and estimated equilibrium index from the Equitest and VR sensory organization test (SOT), respectively, in four of six tested conditions. Additionally, weak to moderate correlations between force plate measurements and the equilibrium index were noted in several of the conditions.

Conclusions: This research demonstrated the plausibility of using VR as an alternative method to conduct the SOT. Ongoing development and testing of virtual environments are necessary before employing the technology as a replacement to current clinical tests.

JMIR Serious Games 2020;8(4):e19580

doi:10.2196/19580

Keywords



The Equitest system (Neurocom) is a computerized dynamic posturography device used by health care providers and clinical researchers to safely test an individual’s postural control. Implementing the sensory organization test (SOT) using the Equitest system requires individuals to process and integrate cues from the visual, vestibular, and proprioceptive systems. This test provides clinicians and researchers with an equilibrium score for each tested condition, a sensory analysis score, a strategy analysis, and a center of gravity (COG) alignment. While the Equitest system has evaluative and rehabilitative value, it may be limited owing to its cost and lack of portability. Moreover, the performance variables provided by the Equitest system are limited, representing gross outcome measures derived only from sagittal plane movement dynamics [1]. Recent advances in technology provide opportunities to reduce these limitations by providing more mobile and cost-effective tools while also observing a wider array of postural characteristics. The purpose of this research was to evaluate the validity of using virtual reality (VR) and a force plate as an alternative to the Equitest system.

The SOT has been the dominant clinical test to assess sensory integration in the context of postural control for neurologic disorders and deficits. With the wide use of clinical dynamic posturography over the last 30 years, the Equitest system has become widely accepted as the gold standard to assess postural stability and balance in several populations (eg, children, aging adults, and military personnel) and clinical groups (eg, those with concussion, vertigo, Parkinson disease, and Alzheimer disease). By systematically disrupting the visual and somatosensory information available to an individual, it is possible to distinguish someone’s reliance on the following three major sensory systems during balance tasks: the visual, somatosensory, and vestibular systems. Conveniently, the Equitest system provides an equilibrium score (indicating how little participants swayed) during each test, as well as a sensory analysis score (indicating how much they relied on each system) and strategy analysis (indicating the hip versus ankle strategy) for the battery of conditions.

While the Equitest system provides a quick evaluative tool for clinicians and researchers, it is not without limitations. First, these outcome measures are derived solely from sagittal plane movements and may not reflect a complete assessment of an individual’s postural control. Second, the costs associated with the Equitest system may limit its availability in underserved communities or during times immediately following an injury (such as a sports concussion). As an alternative to the Equitest system, it may be possible to combine more recent technologies, that is, portable force plates and VR, to ameliorate these drawbacks. When these technologies are combined, they greatly reduce the cost for a clinician to own testing equipment, as well as offer the opportunity to have a portable solution that could be taken into the field. Moreover, portable force plates present the possibility to record and assess a wider range of data, such as medial-lateral dynamics, and customize the outcomes to specific clinical goals. Likewise, VR headsets have continued to improve in quality and decrease in cost, and continued developments may lead to the ability to accurately track movements in VR without additional hardware components such as force plates.

In keeping up with technological advancements, it is important to determine how new technologies can measure up to the “gold standards” they will eventually replace. Currently, VR is approaching this standard and is consistently shown to be a valuable tool to conduct postural and motor control research. Previous research has found no difference between static balance in a physical environment versus a virtual environment [2]. Additionally, several scholars have supported the efficacy of VR for use in balance assessments in a range of clinical populations, such as those with concussion, stroke, Parkinson disease, and high age [3-7]. Continuing in this trend, a large body of research has shown positive results in using VR to enhance training and rehabilitation for balance-related dysfunction [8-11]. Overall, VR has been demonstrated to accurately assess balance in addition to providing a customizable means to enhance clinical outcomes.

The purpose of this research was to compare the Equitest system to a VR balance assessment designed to mimic the SOT in a young healthy population. It was hypothesized that the equilibrium score would demonstrate high limits of agreement between the two testing conditions, supporting VR as a viable option to decrease cost and increase the accessibility of postural assessment techniques. By illustrating the viability of VR to emulate current clinical practices, future progress can focus on improving and optimizing the implementation of VR in clinical standards of care and applications to more populations of interest.


Participants

A convenience sample of 20 college-aged individuals (Table 1) was recruited to participate in this study. All participants were healthy individuals with no prior history of neurological or physical injury or dysfunction. Upon arrival, participants provided informed consent. All procedures were approved by the institutional review board, and no adverse events were encountered.

Table 1. Participant demographics.
CharacteristicMale (n=7), mean (SE)Female (n=13), mean (SE)
Age (years)20.8 (0.4)20.9 (0.37)
Height (m)1.79 (0.03)1.66 (0.02)
Weight (kg)77.4 (5.58)62.8 (3.33)

Experimental Design

After providing informed consent, participants completed a SOT in two blocks, using the Equitest system and using VR. Blocks of tests were counterbalanced, and conditions within blocks were randomized.

During the Equitest SOT, participants wore a harness that supported their weight in case they lost balance. Researchers helped participants into the harness so it fit comfortably and safely. The conditions during the clinical test included (1) eyes open on a stable surface, (2) eyes closed on a stable surface, (3) eyes open with a sway-referenced surround, (4) eyes open on a sway-referenced surface, (5) eyes closed on a sway-referenced surface, and (6) eyes open with both a sway-referenced surround and surface.

In the VR SOT, participants removed any glasses and wore a head-mounted display (HTC Vive, HTC). Participants adjusted the headset to ensure clarity in the virtual environment using a black screen with a textbox. To compare our VR SOT to existing SOT research performed with real machines, we created a virtual scale model of the patterns used inside of the Equitest balance system. We placed this model in the center of a white virtual testing room (10 m × 9 m in size). These models and the testing software were created using Unity 3D (v. 2018.2.10f1; Unity Technologies). Our software allowed us to test users with the following three different types of VR tracking: no tracking, head rotation tracking only, and six degrees of freedom (6DoF) tracking (Figure 1). The “no tracking” option creates an experience where the objects viewed move with the user’s head as if they are attached. The second option, which is common in first-generation VR headsets, such as the Oculus DK1 and Google Daydream, is somewhat natural until users lean in a direction that moves their torso. The last of these most closely mimics reality.

Balance was tested in the following conditions in the VR environment: in a completely dark environment, eyes open in an environment that mimics the clinical test (6DoF tracking), eyes open in an environment that mimics the surround of the clinical test and moves and rotates with the participant’s head (no tracking), and eyes open in an environment that mimics the surround of the clinical test and moves forward and backward with the participant’s head but does not react to head rotation (head tracking only). Each condition was completed on a stable surface and on a foam surface.

For each balance condition, in both the clinical test and the VR test, participants completed two trials of 20 seconds. The order of the trials was counterbalanced between the clinical test and VR blocks, and the order of the conditions was randomized within the clinical test and VR blocks. In total, participants completed 28 trials (six clinical testing conditions × two trials each, four VR testing conditions × two surface conditions × two trials each) of 20 seconds of stationary balance. All participants provided written consent prior to beginning the experimental protocol.

Figure 1. Effect of the head tracking condition in virtual reality on a user's view with translation or rotation of the head. 6DoF: six degrees of freedom.
View this figure

Data Reduction

The Equitest system calculated the equilibrium index (EI) during each SOT condition [12], and it represents the extent to which a participant sways forward or backward within a theoretical limit of 12.5° of displacement. If the participant has no sway, a score of 100 would be received, and if the participant exhibits 12.5° or greater sway (combined forward and backward), a score of 0 would be received. During the VR conditions, participants completed the test on top of a portable force plate (Bertec Inc) that collected center of pressure (COP) data at 50 Hz. Custom MATLAB (Mathworks Inc) scripts were used to detrend and filter the data (20-ms moving average filter) and subsequently calculate the estimated equilibrium index (eEI), 95% ellipse area, path length, and anterior-posterior (AP) detrended fluctuation analysis scaling exponent alpha (DFA α) from the COP data. The 95% ellipse area, path length, and AP DFA α calculations are described elsewhere and represent typical spatiotemporal characteristics of balance [13,14]. The eEI metric was derived based on the EI used by the Equitest system. To simplify this process, the forward and backward sway angles were calculated as the inverse sine function of the anterior and posterior COP displacement, respectively, divided by an estimated COG height (56% of the participant height). The first trial of each condition served as a familiarization period, and only the final trial of each condition was used for analysis. Data that were outside of three times the SD from the mean of its experimental condition were removed from the analysis. In this manner, one trial each from SOT 3 and SOT 4 was removed, along with their VR condition pair.

Statistical Analysis

To assess the relationship between EI and eEI, intraclass correlations of consistency and absolute agreement were calculated for similar conditions (Table 2). Intraclass correlation coefficient (ICC) values were interpreted as poor (<0.5), moderate (0.5-0.75), good (0.5-0.9), and excellent (>0.9) reliability [15]. Additionally, Pearson correlation coefficients were calculated to quantify the extent to which force plate measurements were associated with the EI calculated by the Equitest system within similar conditions. Correlation coefficients were interpreted as negligible (<0.3), weak (0.3-0.5), moderate (0.5-0.7), strong (0.7-0.9), or very strong (>0.9) relationships between pairs of variables [16].

Table 2. Summary of all testing conditions, their abbreviations, and the quality of visual, somatosensory, and vestibular information available in the condition.
Condition abbreviationEquitestVirtual realityInformation qualitya

SOTb 1Eyes opened on a stable surfaceStable virtual surround on a stable surfaceVisc–Somd–Vese
SOT 2Eyes closed on a stable surfaceBlacked out environment on a stable surfaceSom–Ves
SOT 3Eyes opened with a sway-referenced surroundHead-referenced virtual surround on a stable surfaceVis–Som–Ves
SOT 4Eyes opened on a sway-referenced surfaceStable virtual surround on a foam surfaceVis–Som–Ves
SOT 5Eyes closed on a sway-referenced surfaceBlacked out environment on a foam surfaceSom–Ves
SOT 6Eyes opened with a sway-referenced surround and on a sway-referenced surfaceHead-referenced virtual surround on a foam surfaceVisSom–Ves

aIn the column, normal text indicates accurate and italic text indicates inaccurate.

bSOT: sensory organization test.

cVis: visual.

dSom: somatosensory.

eVes: vestibular.


Data Presentation and Assessment of the Raw Data

Boxplots of the data showing the median (thick line), IQR (box edges), and 95% CI (whiskers) for each condition were created (Figure 2). Visual inspection of the data indicated symmetry in most conditions and increased variability in the more challenging conditions (conditions 4-6).

Figure 2. Boxplots of all data collected in comparable SOT (light) and VR (dark) conditions. The median value (thick line), IQR (box edges), and 95% CI (whiskers) are indicated. SOT: sensory organization test; VR: virtual reality.
View this figure

Reliability of the eEI

Intraclass correlations between EI and eEI in similar conditions were evaluated and are presented alongside Bland-Altman plots in Figure 3 [17]. SOT conditions 1, 2, 3, and 6 demonstrated moderate consistency and absolute agreement with their similar VR condition counterparts. Meanwhile, SOT conditions 4 and 5 showed poor consistency and absolute agreement with similar VR conditions. The Bland-Altman plots provide a visual representation of agreement between two measurements by plotting the absolute agreement or mean difference between measurements on the vertical axis against the average of the two measurements on the horizontal axis.

Figure 3. Bland-Altman plots comparing the equilibrium index and estimated equilibrium index from the Equitest and VR SOT, respectively. The Pearson correlation coefficient (r), intraclass correlation coefficient for absolute agreement (ICCa), and intraclass correlation coefficient for consistency (ICCc) are provided. SOT: sensory organization test; VR: virtual reality.
View this figure

Correlation of the EI With Force Plate Measurements

Pearson correlation coefficients were calculated between the Equitest EI and balance measures derived from COP data (Table 3). Weak to moderate significant correlations were identified between EI and eEI in SOT conditions 1 (r=0.454, P=.045), 2 (r=0.566, P=.009), 3 (r=0.652, P=.002), and 6 (r=0.597, P=.005). Additionally, weak to moderate significant correlations were identified between EI and 95% ellipse area in conditions 1 (r=−0.453, P=.045), 2 (r=−0.506, P=.02), and 6 (r=−0.500, P=.03) and AP DFA α in condition 1 (r=−0.511, P=.02). No other relevant correlations were identified between the Equitest EI and balance measurements derived from the COP data.

Table 3. Pearson correlation coefficients between force plate measurements (columns) and the equilibrium index during each sensory organization test condition.
ConditioneEIa95% ellipse areaPath lengthAP DFA αb
SOTc 1




r0.454−0.453−0.130−0.511

P.045.045.59.02
SOT 2




r0.566−0.506−0.400−0.041

P.009.02.08.86
SOT 3




r0.652−0.329−0.068−0.234

P.002.17.78.33
SOT 4




r0.209−0.143−0.332−0.007

P.39.56.16.98
SOT 5




r0.052−0.242−0.2410.027

P.83.30.31.91
SOT 6




r0.597−0.500−0.334−0.174

P.005.03.15.46

aeEI: estimated equilibrium index.

bAP DFA α: anterior-posterior detrended fluctuation analysis scaling exponent alpha.

cSOT: sensory organization test.


This research has demonstrated the plausibility of using VR as an alternative to the Equitest when conducting a SOT. Although not a perfect replacement, eEI demonstrated reasonable correlations and ICCs with the clinical standard in several of the SOT conditions. Continued improvements to the VR testing environment need to be made to have more confidence in its use as a potential replacement. For example, the VR device may do a good job at mimicking the visual conditions of the SOT, but the foam mat might not equivocally disrupt somatosensory information compared with the SOT. This is supported by seeing higher correlations between EI and eEI in the intact than inaccurate somatosensory conditions (conditions 1, 2, and 3 versus conditions 4 and 5). Additionally, this study identified a number of correlations between the Equitest system and typical balance measurements derived from COP data on a force plate. Aside from eEI, 95% ellipse area and AP DFA α had some correlations with the clinical test. It is not surprising that these correlations were somewhat sparse as they distinctly measure different characteristics of balance. The SOT measures only AP sway magnitude, while COP data can be used to calculate sway magnitude in the frontal and sagittal planes combined or to measure aspects of how variability is structured in an individual plane. For example, 95% ellipse area quantifies the gross postural control behavior during quiet stance [18] and AP DFA α quantifies the structure of variability within an individual’s AP sway trajectory (ie, how random or deterministic the data is) [19], whereas EI evaluates how close an individual gets to a theoretical limit of stability [20]. The measures evaluated in this study were selected to represent a small array of postural control measurements, and future research should evaluate the clinical utility of individual metrics.

The recent surge in consumer-ready VR headsets has the potential to greatly reduce the cost of conducting balance assessments while also providing additional accessibility to sites outside of the clinic, for example, on the sideline during an athletic event. Likewise, using force plates opens access to raw, processed, and derived outcome measures that take advantage of the full scope of postural dynamics and present the opportunity to have more accurate information at the clinician’s disposal. In the future, it may even be possible to accurately assess balance (and gait) using only the self-contained tracking of VR headsets. This research serves as a point from which we can merge motor control assessments with the accelerating advancements in consumer technologies.

Conflicts of Interest

None declared.

  1. Chiarovano E, Wang W, Rogers SJ, MacDougall HG, Curthoys IS, de Waele C. Balance in Virtual Reality: Effect of Age and Bilateral Vestibular Loss. Front Neurol 2017;8:5 [FREE Full text] [CrossRef] [Medline]
  2. Robert MT, Ballaz L, Lemay M. The effect of viewing a virtual environment through a head-mounted display on balance. Gait Posture 2016 Jul;48:261-266. [CrossRef] [Medline]
  3. Lloréns R, Noé E, Colomer C, Alcañiz M. Effectiveness, usability, and cost-benefit of a virtual reality-based telerehabilitation program for balance recovery after stroke: a randomized controlled trial. Arch Phys Med Rehabil 2015 Mar;96(3):418-425.e2 [FREE Full text] [CrossRef] [Medline]
  4. Mirelman A, Maidan I, Deutsch JE. Virtual reality and motor imagery: promising tools for assessment and therapy in Parkinson's disease. Mov Disord 2013 Sep 15;28(11):1597-1608 [FREE Full text] [CrossRef] [Medline]
  5. Morel M, Bideau B, Lardy J, Kulpa R. Advantages and limitations of virtual reality for balance assessment and rehabilitation. Neurophysiol Clin 2015 Nov;45(4-5):315-326 [FREE Full text] [CrossRef] [Medline]
  6. Saldana S, Marsh AP, Rejeski WJ, Haberl J, Wu P, Rosenthal S, et al. Assessing balance through the use of a low-cost head-mounted display in older adults: a pilot study. Clin Interv Aging 2017;12:1363-1370 [FREE Full text] [CrossRef] [Medline]
  7. Wright WG, McDevitt J, Tierney R, Haran FJ, Appiah-Kubi KO, Dumont A. Assessing subacute mild traumatic brain injury with a portable virtual reality balance device. Disabil Rehabil 2017 Jul;39(15):1564-1572 [FREE Full text] [CrossRef] [Medline]
  8. de Rooij IJ, van de Port IG, Meijer JW. Effect of Virtual Reality Training on Balance and Gait Ability in Patients With Stroke: Systematic Review and Meta-Analysis. Phys Ther 2016 Dec;96(12):1905-1918 [FREE Full text] [CrossRef] [Medline]
  9. Meyns P, Pans L, Plasmans K, Heyrman L, Desloovere K, Molenaers G. The Effect of Additional Virtual Reality Training on Balance in Children with Cerebral Palsy after Lower Limb Surgery: A Feasibility Study. Games Health J 2017 Feb;6(1):39-48 [FREE Full text] [CrossRef] [Medline]
  10. Villiger M, Liviero J, Awai L, Stoop R, Pyk P, Clijsen R, et al. Home-Based Virtual Reality-Augmented Training Improves Lower Limb Muscle Strength, Balance, and Functional Mobility following Chronic Incomplete Spinal Cord Injury. Front Neurol 2017;8:635 [FREE Full text] [CrossRef] [Medline]
  11. Yen CY, Lin KH, Hu MH, Wu RM, Lu TW, Lin CH. Effects of virtual reality-augmented balance training on sensory organization and attentional demand for postural control in people with Parkinson disease: a randomized controlled trial. Phys Ther 2011 Jun;91(6):862-874 [FREE Full text] [CrossRef] [Medline]
  12. Nashner LM. EquiTest system operator’s manual, version 4.04. Clackamas, OR, USA: NeuroCom International, Inc; 1992.
  13. Duarte M, Freitas SM. Revision of posturography based on force plate for balance evaluation. Rev Bras Fisioter 2010;14(3):183-192. [Medline]
  14. Peng CK, Havlin S, Stanley HE, Goldberger AL. Quantification of scaling exponents and crossover phenomena in nonstationary heartbeat time series. Chaos 1995;5(1):82-87. [CrossRef] [Medline]
  15. Koo TK, Li MY. A Guideline of Selecting and Reporting Intraclass Correlation Coefficients for Reliability Research. Journal of Chiropractic Medicine 2016 Jun;15(2):155-163 [FREE Full text] [CrossRef]
  16. Mukaka MM. Statistics corner: A guide to appropriate use of correlation coefficient in medical research. Malawi Med J 2012 Sep;24(3):69-71 [FREE Full text] [Medline]
  17. Bland J, Altman D. Measuring agreement in method comparison studies. Stat Methods Med Res 1999 Jun 01;8(2):135-160 [FREE Full text] [CrossRef]
  18. Strang AJ, Haworth J, Hieronymus M, Walsh M, Smart LJ. Structural changes in postural sway lend insight into effects of balance training, vision, and support surface on postural control in a healthy population. Eur J Appl Physiol 2011 Jul;111(7):1485-1495 [FREE Full text] [CrossRef] [Medline]
  19. Rhea CK, Silver TA, Hong SL, Ryu JH, Studenka BE, Hughes CM, et al. Noise and complexity in human postural control: interpreting the different estimations of entropy. PLoS One 2011 Mar 17;6(3):e17696 [FREE Full text] [CrossRef] [Medline]
  20. Chaudhry H, Findley T, Quigley KS, Bukiet B, Ji Z, Sims T, et al. Measures of postural stability. J Rehabil Res Dev 2004 Sep;41(5):713-720 [FREE Full text] [CrossRef] [Medline]


6DoF: six degrees of freedom
AP: anterior-posterior
COG: center of gravity
COP: center of pressure
DFA α: detrended fluctuation analysis scaling exponent alpha
eEI: estimated equilibrium index
EI: equilibrium index
ICC: intraclass correlation coefficient
SOT: sensory organization test
VR: virtual reality


Edited by N Zary; submitted 23.04.20; peer-reviewed by PC Wang; comments to author 04.07.20; revised version received 10.09.20; accepted 13.11.20; published 16.12.20

Copyright

©Matthew William Wittstein, Anthony Crider, Samantha Mastrocola, Mariana Guerena Gonzalez. Originally published in JMIR Serious Games (http://games.jmir.org), 16.12.2020.

This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR Serious Games, is properly cited. The complete bibliographic information, a link to the original publication on http://games.jmir.org, as well as this copyright and license information must be included.